This procedure will walk you through manually restoring your Proteus's Brake Coefficients - to ensure resistance is as accurate as possible.
- After startup, go through sensor orientation, then navigate to the "Configure Proteus" page. from the bottom bar of the page, select "Calibration".
- Choose "Expert Calibration" mode.
- In the middle you'll see "Brake Coefficients".
- If your coefficients display red and match those in the image below the machine has reset to its default coefficients and needs to be adjusted. If you do not see red coefficients the coefficients are correct.
- On the left, tap the "Load Calibration Sets" button.
- Select the bottom-most set on the list, and click "Show".
- Click "Apply to Proteus" to restore your unit's coefficients.
- Ensure that the coefficients are no longer red. If not, the unit is ready to use.